Specifications
Platform | Optima AUC |
Maximum Speed | 60,000 rpm |
Maximum g-Force | ||289,848.00|x g |
Acceleration | 5-400 rpm/second (above 500 rpm), 400 max acceleration |
Deceleration | Maximum, Slow & Off |
Speed Control | ± actual rotor speed, ± 2 rpm at steady state |
Timed Run | 1 to 999 hrs, 59 mins (including hold runs) |
Max Capacity Volume Range | 0 - 0.4 mL |
Speed Display | 1,000 to rated speed in 100 rpm increments |
Temperature Range(s) | Ambient temperature range: 10 to 30°C, Set temp: 7-40° C (1°C increments at > 25-30°C ambient temp); 4-40°C (1°C increments at <25°C ambient temp) |
Rotor ID | AN-60 |
Zonal Capability | No |
Pollution Degree | 2 |
Vacuum and Pump | Below 5 microns (0.7 Pa) |
Operating System(s) | LINUX |
Software | Experimental portal can be ran on MS IE 10 (or above); Safari 8 (or above); Chrome 42 (or above) |
Display | 15.3" high-resolution color touch screen |
Operating Mode(s) | LINUX for remote monitoring; GUI (any OS) |
Finish | Rotor pad: Coated polycarbonate, Top & front surfaces: Polyurethane, Other surfaces: Kydex plastic & polyurethane painted aluminum |
Weight | 508.5 kg / 1121 lb |
Width | 94 cm / 37 in |
Depth | 68.1 cm / 26.8 in |
Height | 125.7 cm / 49.5 in |
Clearances | 5.1 cm (sides) & 15.2 cm (rear) |
Humidity Restrictions | <80% at <30°C (non-condensing) |
Power Requirements | 30 A branch circuit, 50/60 Hz, 200-240 VAC |
NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS
The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.